What is the Difference Between Detection Limit and Quantitation Limit? — How to Handle, Report, and Include "Not Detected" in Capability Analysis

By: QTank Published: 11/16/2026 Views: 91
Current rating: ★★★☆☆ Rate this Equivalent to 8 ratings

1. Applicable Scenarios

In a nutshell: The detection limit (LOD) answers "Can it be detected?", the quantitation limit (LOQ) answers "Is this number reliable?", and the reporting limit (RL) answers "How do we communicate externally?" The consequences of mixing these three values are very specific—filling control charts with "Not Detected" as 0, or reporting unreliable numbers to customers as compliance evidence.

Value Question Answered Common Calculation Method Permitted Usage
Detection Limit (LOD) Is it present, can it be seen? 3×s (blank) or 3.3×s₀/slope Only to determine "Not Detected," not to report numerical values
Quantitation Limit (LOQ) Is this number reliable? 10×s (blank) or 10×s₀/slope Below this, report "<LOQ," do not include in mean calculations
Reporting Limit (RL) How to communicate externally Defined by the company, ≥LOQ, rounded to a standard value The sole criterion for report templates and system fields

Applicable: Results that frequently fall near the limit values or below the method's capability—RoHS/ELV harmful substances, migration amounts in food contact materials, VOCs and odors, cleaning residues, coating heavy metals, trace moisture, cleanliness particle counts, and microorganisms. Industries include electronics, automotive materials, medical device cleaning validation, food contact, and toys. In terms of scale, any company with its own laboratory or that regularly commissions third-party testing cannot avoid these three values. The timing is crucial: when introducing a new method, changing instruments, replacing chromatography columns or reagent batches, adding new material grades, tightening customer limits, or during audits requiring method capability evidence (ISO/IEC 17025, IATF 16949 external laboratory qualification).

Not Applicable: Indicators with routine values already ten times higher than the LOQ, such as dimensions, weight, and thickness—these should follow GR&R and measurement uncertainty, calculating LOD is unnecessary. For indicators where the purpose is not to determine "Is it present/exceeding?" but "Is the process stable?"—use process capability. For fully inspected items where actual values are far from the limits, LOD is irrelevant to release decisions.

A practical entry criterion: When "Regulatory or customer limits < 10 times the current method's quantitation capability" or when "Not Detected" appears in reports, LOD/LOQ assessment is necessary. Otherwise, conducting these assessments is not only unnecessary but can also be misleading.

2. Criteria and Thresholds

1. Sample Size: Independent repetitions of blank or spiked blank, n≥10. GB/T 27417-2017 clearly states "No fewer than 10 independent tests." Why can't it be skipped: The standard deviation itself is an estimate, and its relative standard error is approximately 1/√(2(n−1))—for n=3, it's 50%; for n=7, it's 29%; for n=10, it's 24%; for n=20, it's 16%. LOD calculated from 3 blank tests can vary by a factor of two with a new batch of reagents: s changes from 0.08 to 0.15, and LOD changes from 0.24 to 0.45, making the "method capability" inconsistent.

2. Two calculation methods, choose one and document it in the method file.

  • Blank Standard Deviation Method (preferred): LOD = 3×s, LOQ = 10×s, where s is the standard deviation of n≥10 blank results.
  • Calibration Curve Method (when blanks cannot be produced at a low level): LOD = 3.3×s₀/slope, LOQ = 10×s₀/slope, where s₀ is the standard deviation of residuals at low concentration points. The calibration curve must have at least 6 points (including blank), with a correlation coefficient r≥0.99 for quantitative methods and r≥0.98 for screening methods.

3. Signal-to-Noise Ratio (S/N) can only prove instrument performance, not method capability. S/N≥3 is the instrument noise level, without matrix or pre-treatment loss. GB/T 27417-2017 is very clear: S/N can be used to evaluate instrument performance but is not suitable for assessing method detection limits. Companies claiming method LOD based on S/N=3 is the easiest to challenge during audits.

4. Method Capability Acceptance Form (do not issue reports if not met).

Check Item Threshold What to Do if Not Met
RSD at LOQ for 6 Repetitions ≤10% (10%~15% must be noted in the report) Increase LOQ by one standard value or modify pre-treatment
Spike Recovery Rate at LOQ 80%~120% (trace amounts can be relaxed to 70%~130% with explanation) Investigate matrix effects, internal standards, and digestion losses
Calibration Linearity r≥0.99 (quantitative) / r≥0.98 (screening) Reduce concentration range or modify weighted fitting
Relationship Between LOQ and Limits LOQ≤limit/10 can determine "compliance with margin"; limit/10<LOQ≤limit/3 can only determine "exceedance"; LOQ>limit/3 cannot be used for release Change method (e.g., from colorimetry to ICP-MS) or modify decision rules
Decision Band Decision limit = limit − U (k=2); RSD 10% at LOQ corresponds to a relative expanded uncertainty of about 20% For a 1 ppm limit, the decision band is approximately ±0.2 ppm

5. Rules for Resolving Discrepancies Among Three Parties. The supplier's report states "ND," the customer requires "<LOQ and LOQ≤1/10 of the limit," and the laboratory's method LOQ only reaches 1/2 of the limit—none of the three are wrong, but the method was not locked in during the quotation stage. The resolution order is: customer/regulatory method capability > supplier's method at factory > company's internal method; if the method capability is not met, the test results can only be used for process reference, not for release. This is why the testing method must be included in the quotation review: the single cost of colorimetry and ICP-MS can differ by 5 to 10 times, and changing methods requires re-validation and customer change notification, with the saved testing fees far from covering the costs.

3. Executable Steps

  1. Calculate "How Low is Needed" and write it in the first line of the method confirmation plan. Target LOQ = limit ÷ 10 (can be relaxed to ÷3 if compliance margin is sufficient). This step sets the criteria, not just what the lab feels is "sensitive enough."
  2. Conduct capability tests based on the matrix. Perform 10 blank tests, and repeat 6 times at 0.5×, 1×, and 2× the target LOQ levels (same person, same instrument, same day), recording s, recovery rate, and RSD. Additionally, conduct intermediate precision tests over 3 days × 2 times. Data from a single day cannot represent daily operations.
  3. Determine LOQ and Reporting Limit (RL). Set LOQ at the lowest concentration where "RSD≤10% and recovery rate 80%~120%," rounding up to a standard value (e.g., 0.5, 1.0, 2.0 ppm). RL should not be less than LOQ and should align with the customer's report format and system fields.
  4. Fix the Reporting and Decision Criteria. For results ≥LOQ: report the numerical value with U; for LOD≤results<LOQ: report "<LOQ" with the LOQ value in parentheses; for results<LOD: report "Not Detected." Use a consistent decision limit of "limit − U." Document this in report templates and system fields, not relying on human memory.
  5. Set Re-verification Triggers and Routine Controls. Re-verify 6 times at LOQ when changing columns, reagent batches, major instrument repairs, new personnel onboarding, or new matrices. Perform at least one re-verification annually if no changes occur. Include blanks and spiked samples (spike amount 1~2×LOQ) in routine controls.
  6. Review Censoring Proportion Before Capability Analysis. If the "Not Detected" proportion is <50%, use RL for conservative estimation and clearly state "substituted value = RL" in the report; if ≥50%, mean and standard deviation cannot be estimated, and use the rule "all <RL is qualified," do not output Cpk.

4. Counterexamples and Misconceptions

  1. Treating "Not Detected" as 0 in SPC. The lower limit of the control chart is drawn below zero, with points clustering around 0, triggering a false out-of-control signal "7 consecutive points on one side of the center line." When calculating Cpk, both the mean and standard deviation are distorted, making the direction of the distortion unpredictable—using the same data, substituting 0 and substituting RL can lead to opposite conclusions.
  2. Calculating LOD from 3 Blank Tests. A single accidental blank fluctuation determines the method capability commitment, which does not match when the customer re-tests or during audit comparisons, rendering the entire method confirmation document invalid.
  3. Using S/N=3 as the Method Detection Limit. It only proves that the instrument can detect, not that the method can still detect after pre-treatment, matrix, and digestion losses. Testing matrix-containing samples results in a recovery rate of 60%, immediately invalidating the LOD commitment.
  4. Reporting Numerical Values Using LOD as LOQ. Reporting 0.3 ppm when the actual LOQ is 1.0 ppm leads to a threefold error in the customer's formula margin or compliance margin, along with a "qualified" conclusion.
  5. Using Old LOQ After Method Changes. Changing chromatography columns, reagent batches, or major instrument repairs without re-verification and continuing to use old capability data leaves no continuous evidence during audits.
  6. Using "All Not Detected" to Prove Process Excellence. When the method LOQ is 1/2 of the limit, "Not Detected" only means it cannot be detected, not that the process is good. This is the most typical misuse of the zero-defect criterion.

5. Self-Check List

  • The method file clearly states the calculation method (blank method or curve method), the number of blank repetitions n, and the calculated LOD/LOQ values and dates.
  • Each LOQ has original records of 6 repetitions of RSD and spike recovery rate for the corresponding matrix, not copied from other methods.
  • LOQ ≤ 1/10 (at least 1/3) of the customer/regulatory limit, and this requirement is included in the quotation and contract review.
  • The report template has clear formats for "≥LOQ / <LOQ / Not Detected" results, with a consistent decision limit of "limit − U."
  • The proportion of "Not Detected" results and substitution rules are documented; do not output capability indices if the censoring proportion is ≥50%.
  • Re-verification triggers (changing columns, reagent batches, major repairs, new matrices, new personnel) are listed in the method file, along with the most recent re-verification record.

Not Detected is not zero: calculate capability first, then set the criteria.